Document Type

Article

Publication Date

4-26-2017

Publication Title

Biomedical Optics Express

Department

Thayer School of Engineering

Abstract

The design of fluorescence imaging instruments for surgical guidance is rapidly evolving, and a key issue is to efficiently capture signals with high ambient room lighting. Here, we introduce a novel time-gated approach to fluorescence imaging synchronizing acquisition to the 120 Hz light of the room, with pulsed LED excitation and gated ICCD detection. It is shown that under bright ambient room light this technique allows for the detection of physiologically relevant nanomolar fluorophore concentrations, and in particular reduces the light fluctuations present from the room lights, making low concentration measurements more reliable. This is particularly relevant for the light bands near 700nm that are more dominated by ambient lights.

DOI

10.1364/BOE.8.002635

Original Citation

Sexton KJ, Zhao Y, Davis SC, Jiang S, Pogue BW. Optimization of fluorescent imaging in the operating room through pulsed acquisition and gating to ambient background cycling. Biomed Opt Express. 2017 Apr 26;8(5):2635-2648. doi: 10.1364/BOE.8.002635. PMID: 28663895; PMCID: PMC5480502.

Share

COinS